Additel ADT286 Multifunction Reference Thermometer Readout w/ Temperature Scanner

$20,120.00

  • Measure and calibrate SPRTs, RTDs, thermistors and thermocouples
  • 1 PPM resistance ratio accuracy (channel 1)
  • 8 1/2-digit DC multimeter
  • Measure up to 82 channels
  • Bluetooth, WIFI, USB& Ethernet (RJ-45) capable
  • Auto temperature control of Additel and other manufacture’s heat sources
  • Built-in automatic temperature control, data collection, and coefficient generation
  • Support for creating custom control of heat sources with RS-232
  • Auto – zero power feature (self heating compensation)
  • 10.1 in. touch screen display
  • Supports fully automated temperature calibrations with data collection and report generation (no software required)
Categories: , SKU: ADT286-TS-PKG-110V

Description

2019 IDEA Award Winner

The Additel 286 Multifunction Reference Thermometer Readout is an industry first! We have combined the capabilities of a high-end reference thermometer with a highly capable data acquisition system and 8.5 digit multimeter. The ADT286 is capable of scanning and recording up to 82 channels at 10 channels per second. Users can easily configure the ADT286 to perform field calibrations and uniformity studies as well as use the unit in the lab as a precision thermometer and 8.5 digit multimeter. Get more for less with this newest game changer from Additel!

ADT 286 Multifunction Reference Thermometer Readout Scanner Modules

If you’re in need of a precision reference thermometer for your laboratory, then look no further than the Additel 286. The base unit comes with two precision readout channels that can be used to measure your SPRT. Need to calibrate RTDs, PRTs, thermistors or thermocouples? Add a scanner module and you now have the ability to measure 10 RTDs, PRTs, or thermistors and 20 thermocouples. Expand up to 82 channels with our unique easy to use scanner modules. Each 20 channel module is outfitted with our proprietary universal terminals with an industries best cold junction capability second to none. Utilize the module docked atop the ADT286, or connect remotely with cables to suit nearly any unique setup/configuration. Additel also has a process module specifically designed to measure process instrumentation like transmitters and switches. This scanner will also supply loop power for the transmitters.

Designed to make your job easier, the ADT286 has a large sensor library supporting 15 TC types, both standard and special limits, 18 different thermal resistors, CVD, ITS-90, and a large variety of standard curves for RTDs and thermistors. The ADT286 is loaded with special applications such as probe calibration, SPRT calibration, chamber mapping and more. And we continue to add applications on a regular basis!

Automatic Temperature Control and Probe Calibration

The Additel 286 Multifunction Reference Thermometer Readout has preinstalled drivers to control Additel and other manufacturer’s heat sources. Simply connect to one or more heat sources via a communication cable, Ethernet or wireless and now it will automatically control to the set point and desired stability. If your heat source is not on the list, you can easily add the driver yourself so you can run automated calibrations with any heat source.

Now combine the heat source control feature with our probe calibration application and you have a very powerful automatic calibration solution. The probe calibration app allows you to automatically setup and run calibration routines with multiple set points and multiple heat sources, collect data, and develop calibration coefficients — all with one device and without the need of software! Simply place a batch of sensors of any mix and type into your heat source, connect it to the ADT286, run the probe calibration app and come back to a completed test. All that is left to do is generate and export all the calibration data. There’s no need to work with complicated software for communication, set up or coefficient generation. There is no reason to have a calibration technician manually monitor the process and record the data. This Multifunction Reference Thermometer Readout will do all the work for you.

ADT286-RS Resistance Standards

Available in 25 and 100 Ohm values, users can enjoy improved resistance ratio performance by easily plugging one of our reference resisters into channel 2 on the new ADT286. Perfect for calibrating your SPRT’s and high end PRT’s.

Additional information

Weight 18.5 lbs
Dimensions 9.8 × 16.5 × 7.9 in

Each Unit Includes

  • Multifunction Reference Thermometer Readout base unit with 1 Process Scanner Module
  • Shorting Block
  • USB Cable
  • 4 Sets of Test Leads (8 pcs)
  • 9026 4-Wire Test Leads
  • Fuse
  • Calibration Certificate
  • CD Manual

Specifications

General Specifications
Specification Description
Voltage 100V Setting 90V to 110V
120V Setting 108V to 132V
220V Setting 198V to 242V
240V Setting 216V to 264V
Frequency 47Hz to 440Hz.Automatically sensed at power-on
Power Consumption 40VA peak (30Watt average)
Temperature Operating : 0°C to 50°C
Full accuracy :18°C to 28°C
Storage : -20°C to 70°C
Warm-up 60 mins for full uncertainty specifications
Relative Humidity (non-condensing) Operating : 0°C to 28°C < 90%
28°C to 40°C < 75%
40°C to 50°C < 50%
Storage : -20°C to 70°C < 95%
Altitude Operating : 2000 m
Storage : 12000 m
Vibration and Shock Complies with MIL-28800F Class 3
Input Protection 50V all functions, ranges and terminals
Communication USB-A , USB-B , RJ45 , WiFi , Bluetooth
Memory 10G – All data stored with time stamps
Localization English , Chinese
Display 10.1 in (256 mm) TFT color display
Size (H x W x D) 9.8 in (250 mm) x 16.5 in (420 mm) x 7.9 in (200 mm)
Weight 18.5 lb ( 8.39 kg)
Other Conformities CE
Warranty 1 year
Measurement Specifications
Specification Conditions: 60 mins Warm-Up TIme / Enviroment Temperature (18-28) °C
The following specifications apply for the front panel, after at least 60 minutes warm-up.
24-hour specifications are relative to calibration standards and assume a controlled electromagnetic environment per EN 61326.
Resistance Ratio Accuracy (Rx/Rs) using External Rs
Range Reference Resistance Ratio (Rx/Rs) 1 Year (23 ± 5) °°Cppm of Reading
100 Ω 25 Ω 2.00-4.00 1.5
1.10-2.00 0.85
0.90-1.10 0.6
0.50-0.90 1.5
0.25-0.50 2.5
400 Ω 100 Ω 2.00-4.00 2
1.10-2.00 0.81
0.90-1.10 0.26
0.50-0.90 0.95
0.25-0.50 1.2
SPRT/PRT Measurement Accuracy using External Rs
SPRT/PRT Type External Reference Resistance Temperature (°C) ° Resistance Ratio (Rx/Rs) 1 Year(23 ± 5) °C ppm of reading Equivalent to Temperature (mK)
PT25 25 Ω -189.3442 0.22 2.5 0.13
-38.8344 0.84 1.5 0.32
0.01 1 0.6 0.15
231.928 1.89 0.85 0.44
419.527 2.57 1.5 1.11
660.323 3.37 1.5 1.58
PT100 100 Ω -189.3442 0.22 1.2 0.07
-38.8344 0.84 0.95 0.2
0.01 1 0.26 0.07
231.928 1.89 0.81 0.42
419.527 2.57 2 1.47
660.323 3.37 2 2.11
[1] The PT25 indicator is based on a nominal resistance of 25 Ω for Rx.
[2] The PT100 indicator is based on a nominal resistance of 100 Ω for Rx.
[3] The uncertainty of external Rs is not included. The user may choose the ADT280-RS-25/100 standard resistor as external Rs, which has an accuracy of 5 ppm at (23±2) °C.
Resistance Accuracy using Internal Rs
Measurement Range Scanning Speed Resolution 24 Hour (23 ±1) °C 90 Days (23 ± 5) °C 1 year (23 ± 5) °C Excitation Currert Temperature Coefficient
( 0~100 ) Ω Slow Speed 0.01 mΩ 3 ppm or 0.2 mΩ 12 ppm or 0.35 mΩ 15 ppm or 0.35 mΩ ±1 mA/±12 V 3 ppm + 0.01 mΩ
Medium Speed 0.01 mΩ 3 ppm or 0.55 mΩ 12 ppm or 0.7 mΩ 15 ppm or 0.7 mΩ
Fast Speed 0.1 mΩ 3.6 ppm or 1.7 mΩ 12.6 ppm or 1.85 mΩ 15.6 ppm or 1.85 mΩ
( 0~400 ) Ω Slow Speed 0.01 mΩ 3 ppm or 0.3 mΩ 12 ppm or 0.4 mΩ 15 ppm or 0.4 mΩ ±1 mA/±12 V 3 ppm + 0.02 mΩ
Medium Speed 0.01 mΩ 3 ppm or 0.7 mΩ 12 ppm or 0.8 mΩ 15 ppm or 0.8 mΩ
Fast Speed 0.1 mΩ 3.6 ppm or 1.9 mΩ 12.6 ppm or 2 mΩ 15.6 ppm or 2 mΩ
( 0~4000 ) Ω Slow Speed 0.1 mΩ 3 ppm or 4 mΩ 12 ppm or 5 mΩ 15 ppm or 5 mΩ ±0.1 mA/±12 V 3 ppm + 0.2 mΩ
Medium Speed 0.1 mΩ 3 ppm or 8 mΩ 12 ppm or 9 mΩ 15 ppm or 9 mΩ
Fast Speed 1 mΩ 3.6 ppm or 20 mΩ 12.6 ppm or 21 mΩ 15.6 ppm or 21 mΩ
[1] Accuracy Index: ± (ppm of reading or xxmΩ , whichever is greater).
[2] Temperature coefficient index: exceeds (18-28) °C range, increase per degree (ppm reading +xxmΩ).
[3] Specifications are for 4-wire function. For 3-wire, add 0.005 Ω for internal resistance mismatch. For 2-wire, add 0.005 Ω for internal resistance.
[4] Automatic current reversal.
PRT Measurement Accuracy using Internal Rs
Scanning Speed Temperature 24 Hour / °C (23 ±1) °C 90 Days / °C(23 ± 5) °C 1 year / °C(23 ± 5) °C Temperature Coefficient °C/°C
Slow Speed -200 °C 0.0005 0.0008 0.0008 0.0002
0 °C 0.0008 0.0031 0.0038 0.0008
300 °C 0.0018 0.0089 0.0089 0.0018
600 °C 0.0029 0.0146 0.0146 0.003
Medium Speed -200 °C 0.0013 0.0016 0.0016 0.0002
0 °C 0.0014 0.0031 0.0038 0.0008
300 °C 0.002 0.0089 0.0089 0.0018
600 °C 0.0029 0.0146 0.0146 0.003
Fast Speed -200 °C 0.0039 0.0043 0.0043 0.0006
0 °C 0.0044 0.0047 0.0047 0.0013
300 °C 0.0053 0.0093 0.0093 0.0024
600 °C 0.0059 0.0152 0.0152 0.0036
[1] The indicator is based on the electrical accuracy of the 4-wire PT100 PRT and does not include the accuracy of the PRT itself.
[2] Temperature maximum Resolution is 0.0001 °C.
Thermocouple Voltage Accuracy
Test Range Scanning Speed Resolution 24 hours (23 ±1) °C 90 days (23 ±5) °C 1 year (23 ±5) °C Input Resistance Temperature Coefficient
(-100-100) mV Slow Speed 0.01 µV 5 ppm + 2 ppm 10 ppm + 4 ppm 14 ppm + 4 ppm 10 MΩ or >10 GΩ 1 ppm + 0.1 µV
Medium Speed 0.01 µV 5 ppm + 6 ppm 10 ppm + 8 ppm 14 ppm + 8 ppm
Fast Speed 0.1 µV 5 ppm + 22 ppm 10 ppm + 24 ppm 14 ppm + 24 ppm
[1] Accuracy Index: ± (ppm of reading + ppm of FS).
[2] Temperature coefficient index: Exceed the range of (18-28) °C, increase (ppm of reading + xxμV)/°C.
Thermocouple Cold Junction Accuracy
CJC Accuracy ±0.1 °C , 1 year, 23 °C ± 5°C
Environmental Coefficient Beyond (18 ~ 28) °C, add 0.02 °C / °C
Other Each signal scanner has 10 cold Junction temperature sensors
Thermocouple Temperature Accuracy
Type Temp. 24 Hour/°C 90 days/°C 1 year /°C
(23 ±1) °C (23 ± 5) °C (23 ±5) °C
Fast Speed Medium Speed Slow Speed Fast Speed Medium Speed Slow Speed Fast Speed Medium Speed Slow Speed
E -200 0.089 0.038 0.022 0.099 0.047 0.031 0.1 0.049 0.033
-100 0.049 0.021 0.012 0.054 0.026 0.017 0.055 0.026 0.017
-40 0.041 0.017 0.009 0.045 0.021 0.013 0.045 0.021 0.014
0 0.038 0.015 0.009 0.041 0.019 0.012 0.041 0.019 0.012
155 0.031 0.013 0.008 0.035 0.017 0.011 0.036 0.017 0.012
350 0.029 0.013 0.008 0.033 0.017 0.012 0.035 0.018 0.013
660 0.031 0.014 0.009 0.036 0.02 0.015 0.039 0.022 0.017
1000 0.034 0.017 0.012 0.042 0.025 0.019 0.046 0.029 0.024
J -200 0.102 0.043 0.025 0.113 0.054 0.036 0.115 0.055 0.037
-100 0.054 0.022 0.013 0.06 0.028 0.018 0.06 0.028 0.019
-40 0.047 0.019 0.011 0.051 0.024 0.015 0.051 0.024 0.015
0 0.044 0.018 0.01 0.048 0.022 0.014 0.048 0.022 0.014
155 0.041 0.017 0.01 0.045 0.021 0.014 0.046 0.022 0.015
350 0.042 0.018 0.011 0.047 0.023 0.016 0.048 0.025 0.018
660 0.039 0.018 0.011 0.046 0.024 0.018 0.048 0.027 0.02
1200 0.044 0.022 0.015 0.054 0.031 0.024 0.059 0.036 0.029
K -200 0.146 0.061 0.035 0.161 0.076 0.05 0.163 0.077 0.051
-100 0.073 0.03 0.017 0.08 0.037 0.024 0.08 0.038 0.025
-40 0.06 0.025 0.014 0.066 0.03 0.02 0.066 0.031 0.02
0 0.056 0.023 0.013 0.061 0.028 0.018 0.061 0.028 0.018
155 0.056 0.023 0.013 0.061 0.029 0.019 0.062 0.03 0.02
350 0.054 0.023 0.014 0.061 0.03 0.02 0.062 0.031 0.021
660 0.055 0.025 0.015 0.063 0.033 0.023 0.066 0.035 0.026
1372 0.073 0.035 0.023 0.087 0.049 0.037 0.093 0.055 0.043
T -200 0.142 0.059 0.034 0.156 0.073 0.048 0.157 0.075 0.049
-100 0.078 0.032 0.018 0.086 0.04 0.026 0.086 0.04 0.026
-40 0.063 0.026 0.015 0.069 0.032 0.02 0.069 0.032 0.021
0 0.057 0.023 0.013 0.062 0.028 0.018 0.062 0.028 0.018
155 0.044 0.019 0.011 0.049 0.023 0.015 0.049 0.024 0.016
350 0.038 0.016 0.01 0.043 0.021 0.015 0.044 0.022 0.016
400 0.037 0.016 0.01 0.042 0.021 0.015 0.044 0.023 0.016
R -40 0.543 0.222 0.124 0.593 0.272 0.173 0.593 0.272 0.173
0 0.416 0.17 0.095 0.454 0.208 0.132 0.454 0.208 0.132
155 0.266 0.109 0.061 0.29 0.134 0.086 0.291 0.134 0.086
350 0.22 0.091 0.051 0.241 0.112 0.072 0.242 0.113 0.073
660 0.192 0.08 0.046 0.212 0.1 0.066 0.214 0.102 0.068
1768 0.188 0.082 0.049 0.213 0.107 0.074 0.219 0.114 0.081
S -40 0.515 0.211 0.117 0.562 0.258 0.164 0.562 0.258 0.164
0 0.407 0.167 0.093 0.444 0.204 0.13 0.444 0.204 0.13
155 0.275 0.113 0.063 0.3 0.138 0.089 0.301 0.139 0.089
350 0.236 0.098 0.055 0.259 0.12 0.078 0.26 0.122 0.079
660 0.214 0.089 0.051 0.236 0.111 0.073 0.239 0.114 0.075
1768 0.222 0.096 0.057 0.25 0.124 0.086 0.257 0.132 0.093
B 250 0.872 0.357 0.199 0.952 0.437 0.278 0.952 0.437 0.279
350 0.619 0.254 0.141 0.676 0.311 0.198 0.676 0.311 0.199
660 0.342 0.141 0.079 0.374 0.173 0.111 0.375 0.175 0.113
1820 0.199 0.085 0.05 0.222 0.108 0.073 0.227 0.113 0.078
N -200 0.224 0.093 0.052 0.246 0.115 0.075 0.247 0.116 0.076
-100 0.106 0.044 0.024 0.116 0.054 0.035 0.116 0.054 0.035
-40 0.089 0.036 0.02 0.097 0.045 0.029 0.097 0.045 0.029
0 0.084 0.035 0.019 0.092 0.042 0.027 0.092 0.042 0.027
155 0.07 0.029 0.017 0.077 0.036 0.024 0.078 0.037 0.024
350 0.062 0.026 0.015 0.069 0.033 0.022 0.07 0.035 0.024
660 0.059 0.026 0.016 0.067 0.034 0.024 0.069 0.036 0.026
800 0.06 0.027 0.016 0.068 0.035 0.025 0.071 0.038 0.028
1000 0.062 0.028 0.018 0.072 0.038 0.028 0.075 0.042 0.031
1200 0.065 0.03 0.019 0.076 0.041 0.031 0.081 0.046 0.035
1300 0.068 0.032 0.02 0.08 0.044 0.033 0.085 0.049 0.038
L -200 0.069 0.029 0.017 0.076 0.036 0.024 0.077 0.037 0.025
-100 0.053 0.022 0.013 0.059 0.028 0.018 0.059 0.028 0.018
-40 0.045 0.019 0.01 0.049 0.023 0.015 0.05 0.023 0.015
0 0.043 0.018 0.01 0.047 0.021 0.014 0.047 0.021 0.014
155 0.04 0.017 0.01 0.044 0.021 0.014 0.045 0.022 0.015
350 0.041 0.018 0.011 0.046 0.023 0.016 0.047 0.024 0.017
660 0.039 0.018 0.011 0.046 0.024 0.018 0.048 0.027 0.02
900 0.035 0.017 0.011 0.042 0.023 0.017 0.045 0.026 0.021
U -80 0.072 0.03 0.017 0.079 0.037 0.024 0.079 0.037 0.024
-40 0.062 0.026 0.014 0.068 0.031 0.02 0.068 0.032 0.02
0 0.056 0.023 0.013 0.061 0.028 0.018 0.061 0.028 0.018
155 0.045 0.019 0.011 0.049 0.023 0.015 0.05 0.024 0.016
350 0.037 0.016 0.01 0.042 0.021 0.014 0.043 0.022 0.016
600 0.034 0.015 0.01 0.039 0.021 0.015 0.041 0.023 0.017
[1] The index is based on the accuracy of the thermocouple electrical measurement of temperature scanner module, does not include the accuracy of the thermocouple itself and the fixed cold junction compensation at 0 °C.
[2] The highest temperature resolution is 0.0001 °C.
Thermistor Accuracy
Measurement Range Scanning Speed Resolution 24 Hour (23 ±1) °°C 90 Days (23 ± 5) °C 1 year (23 ± 5) °°C Excitation Currert Temperature Coefficient
( 0~12 ) kΩ Slow Speed 1 mΩ 10 ppm or 60 mΩ 30 ppm or 80 mΩ 40 ppm or 80 mΩ 10 µA 5 ppm + 10 mΩ
Medium Speed 1 mΩ 10 ppm or 110 mΩ 30 ppm or 130 mΩ 40 ppm or 130 mΩ
Fast Speed 10 mΩ 10 ppm or 210 mΩ 30 ppm or 230 mΩ 40 ppm or 230 mΩ
( 10~120 ) kΩ Slow Speed 10 mΩ 10 ppm 30 ppm 40 ppm 10 µA 5 ppm + 20 mΩ
Medium Speed 10 mΩ 10 ppm + 80 mΩ 30 ppm + 80 mΩ 40 ppm + 80 mΩ
Fast Speed 100 mΩ 10.6 ppm + 200 mΩ 30.6 ppm + 200 mΩ 40.6 ppm + 200 mΩ
( 100~1000 ) kΩ Slow Speed 0.1 Ω 50 ppm 80 ppm 100 ppm 1 µA 5 ppm + 1 Ω
Medium Speed 0.1 Ω 50 ppm + 1 Ω 80 ppm + 1 Ω 100 ppm + 1 Ω
Fast Speed 1 Ω 51 ppm + 2 Ω 81 ppm + 2 Ω 101 ppm + 2 Ω
[1] Accuracy Index: ± (ppm of reading or xxmΩ , whichever is greater).
[2] Temperature coefficient index: exceeds (18-28) °C range, increase (ppm reading +xxmΩ) / °C.
[3] Specifications are for 4-wire function.
Thermistor Temperature Accuracy
Type Scanning Speed Temperature 24 Hour / °C (23 ±1) °C 90 Days / °C(23 ± 5) °C 1 year / °C(23 ± 5) °°C
10 kΩ Slow Speed -40 °C 0.0007 0.0011 0.0014
0 °C 0.0002 0.0006 0.0008
50 °C 0.0004 0.0008 0.0011
100 °C 0.003 0.0039 0.0039
150 °C 0.013 0.0174 0.0174
Medium Speed -40 °C 0.0007 0.0011 0.0014
0 °C 0.0002 0.0006 0.0008
50 °C 0.0008 0.001 0.0011
100 °C 0.0054 0.0064 0.0064
150 °C 0.0239 0.0282 0.0282
Fast Speed -40 °C 0.0007 0.0011 0.0014
0 °C 0.0002 0.0006 0.0008
50 °C 0.0016 0.0016 0.0016
100 °C 0.0104 0.0104 0.0104
150 °C 0.0456 0.0456 0.0456
[1] The indicator is based on the electrical accuracy of the 4-wire thermistor and does not include the accuracy of the thermistor itself.
[2] Temperature maximum Resolution is 0.0001 °C.
DC Voltage Accuracy
Test Range Scanning Speed Resolution 24 hours (23 ±1)°C 90 days (23 ±5)°C 1 year (23 ±5)°C Input Resistance Temperature Coefficient
(-100-100) mV Slow Speed 0.01 µV 5 ppm + 2 ppm 10 ppm + 4 ppm 14 ppm + 4 ppm >10 GΩ or 10 MΩ 1 ppm + 0.1 µV
Medium Speed 0.01 µV 5 ppm + 6 ppm 10 ppm + 8 ppm 14 ppm + 8 ppm
Fast Speed 0.1 µV 5 ppm + 22 ppm 10 ppm + 24 ppm 14 ppm + 24 ppm
(-1-1) V Slow Speed 0.1 µV 2 ppm + 0.3 ppm 8 ppm + 0.6 ppm 14 ppm + 0.6 ppm >10 GΩ or 10 MΩ 1 ppm + 0.2 µV
Medium Speed 0.1 µV 2 ppm + 1.3 ppm 8 ppm + 1.6 ppm 14 ppm + 1.6 ppm
Fast Speed 1 µV 2.6 ppm + 3.3 ppm 8.6 ppm + 3.6 ppm 14.6 ppm + 3.6 ppm
(-10-10) V Slow Speed 1 µV 2 ppm + 0.05 ppm 8 ppm + 0.08 ppm 14 ppm + 0.08 ppm >10 GΩ or 10 MΩ 1 ppm + 0.3 µV
Medium Speed 1 µV 2 ppm + 0.35 ppm 8 ppm + 0.38 ppm 14 ppm + 0.38 ppm
Fast Speed 10 µV 2.6 ppm + 1.05 ppm 8.6 ppm + 1.08 ppm 14.6 ppm + 1.08 ppm
(-50-50) V Slow Speed 10 µV 8 ppm + 1 ppm 32 ppm + 1 ppm 38 ppm + 1 ppm 10 MΩ 5 ppm + 5 µV
Medium Speed 10 µV 8 ppm + 2 ppm 32 ppm + 2 ppm 38 ppm + 2 ppm
Fast Speed 100 µV 8.6 ppm + 7 ppm 32.6 ppm + 7 ppm 38.6 ppm + 7 ppm
[1] Accuracy Index: ± (ppm of reading + ppm of FS).
[2] Temperature Coefficient index: Exceed the range of (18-28) °C, increase (ppm reading + xxμV)/°C.
[3] Any range, the maximum input voltage is 50 V.
DC Current Accuracy
Test Range Scanning Speed Resolution 24 hours (23 ±1) °C 90 days (23 ±5) °C 1 year (23 ±5) °C Burden Voltage Temperature Coefficient
(-100-100) µA Slow Speed 0.01 nA 15 ppm + 3 ppm 50 ppm + 6 ppm 60 ppm + 6 ppm <1 mV 8 ppm + 0.1 nA
Medium Speed 0.01 nA 15 ppm + 7 ppm 50 ppm + 10 ppm 60 ppm + 10 ppm
Fast Speed 0.1 nA 15 ppm + 23 ppm 50 ppm +26 ppm 60 ppm + 26 ppm
(-1-1) mA Slow Speed 0.1 nA 15 ppm + 0.6 ppm 50 ppm +1 ppm 60 ppm + 1 ppm <1 mV 8 ppm + 0.5 nA
Medium Speed 0.1 nA 15 ppm + 1.6 ppm 50 ppm + 2 ppm 60 ppm + 2 ppm
Fast Speed 1 nA 15.6 ppm + 3.6ppm 50.6 ppm + 4 ppm 60.6 ppm + 4 ppm
(-10-10) mA Slow Speed 1 nA 30 ppm + 3 ppm 75 ppm + 6 ppm 80 ppm + 6 ppm <1 mV 8 ppm + 10 nA
Medium Speed 1 nA 30 ppm + 7 ppm 75 ppm + 10 ppm 80 ppm + 10 ppm
Fast Speed 10 nA 30 ppm + 23 ppm 75 ppm + 26 ppm 80 ppm + 26 ppm
(-100-100) mA Slow Speed 10 nA 40 ppm + 0.6 ppm 75 ppm + 1 ppm 80 ppm + 1 ppm <1 mV 8 ppm + 50 nA
Medium Speed 10 nA 40 ppm + 1.6 ppm 75 ppm + 2 ppm 80 ppm + 2 ppm
Fast Speed 100 nA 40.6 ppm + 3.6 ppm 75.6 ppm + 4 ppm 80.6 ppm + 4 ppm
[1] Accuracy Index: ± (ppm of reading + ppm of FS).
[2] Temperature Coefficient index:Exceed the range of range of (18-28) °C, increase (ppm reading+ xxnA)/°C.
[3] Input Protection 0.3A/600V Resettable PTC.
DC Resistance Accuracy
Test Range Scanning Speed Resolution 24 hours (23 ±1) °C 90 days (23 ±5) °C 1 year (23 ±5) °C Excitation Current Temperature Coefficient
(0-100) Ω Slow Speed 0.01 mΩ 3 ppm + 1 ppm 13 ppm + 1.5 ppm 16 ppm + 1.5 ppm 1 mA 3 ppm + 0.01 mΩ
Medium Speed 0.01 mΩ 3 ppm + 5 ppm 13 ppm + 5.5 ppm 16 ppm + 5.5 ppm
Fast Speed 0.1 mΩ 3 ppm + 21 ppm 13 ppm + 21.5 ppm 16 ppm + 21.5 ppm
(0-1) kΩ Slow Speed 0.1 mΩ 3 ppm + 0.2 ppm 12 ppm + 0.3 ppm 15 ppm + 0.3 ppm 1 mA 3 ppm + 0.02 mΩ
Medium Speed 0.1 mΩ 3 ppm + 1.2 ppm 12 ppm + 1.3 ppm 15 ppm + 1.3 ppm
Fast Speed 1 mΩ 3.6 ppm + 3.2 ppm 12.6 ppm + 3.3 ppm 15.6 ppm + 3.3 ppm
(0-10) kΩ Slow Speed 1 mΩ 3 ppm + 0.3 ppm 12 ppm + 0.4 ppm 15 ppm + 0.4 ppm 0.1 mA 3 ppm + 0.2 mΩ
Medium Speed 1 mΩ 3 ppm + 1.3 ppm 12.6 ppm + 1.3 ppm 15 ppm + 1.3 ppm
Fast Speed 10 mΩ 3.6 ppm + 3.3 ppm 12.6 ppm + 3.4 ppm 15.6 ppm + 3.4 ppm
(0-100) kΩ Slow Speed 10 mΩ 3 ppm + 0.2 ppm 12 ppm + 0.3 ppm 15 ppm + 0.3 ppm 0.1 mA 3 ppm + 20 mΩ
Medium Speed 10 mΩ 3 ppm + 0.5 ppm 12 ppm + 0.6 ppm 15 ppm + 0.6 ppm
Fast Speed 100 mΩ 3.6 ppm + 1.3 ppm 12.6 ppm + 1.3 ppm 30.6 ppm + 1.3 ppm
(0-1) MΩ Slow Speed 0.1 Ω 10 ppm + 0.6 ppm 30 ppm + 1 ppm 40 ppm + 1 ppm 10 µA 5 ppm + 0.2 Ω
Medium Speed 0.1 Ω 10 ppm + 1.2 ppm 30 ppm + 0.6 ppm 40 ppm + 0.6 ppm
Fast Speed 1 Ω 10 ppm + 2.6 ppm 30 ppm + 3 ppm 40 ppm + 3 ppm
(0-10) MΩ Slow Speed 1 Ω 50 ppm + 0.4 ppm 80 ppm + 1 ppm 100 ppm + 1 ppm 1 µA 10 ppm + 1 Ω
Medium Speed 1 Ω 50 ppm + 1.4 ppm 80 ppm + 2 ppm 100 ppm + 2 ppm
Fast Speed 10 Ω 50 ppm + 4.4 ppm 80 ppm + 5 ppm 100 ppm + 5 ppm
(0-100) MΩ Slow Speed 10 Ω 150 ppm + 1 ppm 400 ppm + 4 ppm 500 ppm + 4 ppm 0.1 µA 50 ppm + 50 Ω
Medium Speed 10 Ω 150 ppm + 6 ppm 400 ppm + 9 ppm 500 ppm + 9 ppm
Fast Speed 100 Ω 150 ppm + 11 ppm 400 ppm + 14 ppm 500 ppm + 14 ppm
[1] Accuracy Index: ± (ppm of reading + ppm of FS).
[2] Temperature Coefficient index:Exceed the range of range of (18-28) °C, increase (ppm reading+ xx Ω)/°C.
[3] The above is a 4-wire measurement index.
[4] When the range is less than or equal to 10 kΩ, the default is automatic current reversal.
[5] Max Lead Resistance(4-wire ohms): 10 Ω per lead for 100 Ω & 1 kΩ ranges; 100 Ω per lead for 10 kΩ &100 kΩ ranges; 1 kΩ per lead on all other ranges.
Secondary Standard PRT Information
Specification AM1760 Series AM1762 Series
Temperature Range -200 °C to 670 °C
Resistance at 0℃ Nominal 100 Ω Nominal 25 Ω
Temperature Coefficient 0.003925 Ω / Ω / °C
Accuracy ±0.007 °C at -196 °C
±0.006 °C at 0.01 °C
±0.015 °C at 420 °C
±0.025 °C at 660 °C
Drift ±0.004 °C at TPW after 100 hours at 661 °C
Short Term Stability ±0.002 °C
Thermal Shock ±0.002 °C after 10 times thermal cycles from minimum to maximum temperatures
Hysteresis N/A
Self-heating 0.0015 °C at 1 mA current
Response Time 9 seconds for 63% response to step change in water moving at 3 feet per second
Measurement Current 0.5 mA or 1 mA
Sensor Length 42 mm
Sensor Location 5 mm from tip
Insulation Resistance >1000 MΩ at room temperature
Sheath Material Inconeltm
Dimension AM1762-12-SP AM1760-12-SP
0.25 in dia X 12 in (6.35 mm X 305 mm) 0.25 in dia X 12 in (6.35 mm X 305 mm)
AM1762-20-SP AM1760-20-SP
0.25 in dia X 20 in (6.35 mm X 500 mm) 0.25 in dia X 20 in (6.35 mm X 500 mm)
External Leads Teflontm – insulated copper wire, 4 leads, 2.5 meters
Handle Dimension 15 mm (OD) x 65 mm (L)
Handle Temperature Range [1] -50 °C to 160 °C -50 °C to 180° C
Calibration NIST traceable calibration w/ data included
[1] Handle temperatures outside the usable will cause damage to the probe.* PRT Information from www.accumac.com.
Resistance Standards
Specification ADT280-RS-25 ADT280-RS-100
Nominal Resistance 25 Ω 100 Ω
Stability 5 ppm/year
Operating Temperature 23 °C ± 2 °C
Temperature Coefficient 0.5 ppm/°C
Size 57 mm x 57 mm x 45 mm
Weight 0.35 lb (160 g)
Excitation Current 1 mA

Technical Resources

Related products

Our Newsletter Is Coming Soon!

Please check back soon while we get these important updates in place.