Description
Idea behind M550 Impedance Calibrator is to replace countless single value LCR standards with one calibrator that can switch between the standards like a CD changer in your car. The calibrator contains stable and temperature independent resistance standards, partial capacitance standards and partial inductance standards, simulated by RC T-type passive network. Calibration values of both complex parameters of partial standards are displayed on large LCD screen in preselected pairs of parameters as well as voltage, current and frequency readouts from built-in test signal meter. OPEN and SHORT reference positions are available as well.
Easy recalibration system offers either full recalibration in all spot frequency points or simplified offset calibration. History of calibration values of partial standards at reference frequency 1 kHz can be easily recalled.
Calibration domain
M550 was designed for calibration of LCR meters in frequency range up to 1 MHz. The philosophy of M550 is based on remote control and automated calibration. For this reason, the calibrator is equipped with RS-232 and GPIB interfaces and of course, supported in CALIBER/WinQbase calibration software.
Specifications
Resistance | 100 mΩ – 100 MΩ, 20 Hz – 1 MHz ± 0.02% / year @ 1 kHz |
Capacitance | 10 pF – 100 uF, 20 Hz – 1 MHz ± 0.05% / year @ 1 kHz |
Inductance | 10 uH – 10 H, 20 Hz – 1 MHz simulated by T-network circuit ± 0.1% / year @ 1 kHz |
Test Signal Meter
AC voltage | 0.2 – 10 Vrms ± 2% / year |
AC current | 1 nA – 500 mA ± 2% / year |
Frequency | 20 Hz – 100 kHz ± 100 ppm / year |